The Regulation on health technology assessment (HTA) establishes a framework for joint clinical assessments (JCAs) of certain medical devices (MDs) and in vitro diagnostic medical devices (IVDs) at EU level.
JCAs provide a scientific analysis of the clinical evidence on a medical product, including its relative clinical effectiveness and relative safety compared with other health technologies. JCAs on MDs and IVDs take place after their certification by the concerned notified bodies.
Which medical devices and in vitro diagnostic medical devices are subject to a JCA?
JCAs may be carried out for:
- class IIb and class III MDs under Regulation (EU) 2017/745 (MDR), for which the relevant medical device expert panel has provided a scientific opinion as part of the clinical evaluation consultation procedure.
- class D IVDs under Regulation (EU) 2017/746 (IVDR), for which the relevant expert panel has provided its views under the applicable procedure.
Not all devices with an expert panel opinion or view are subject to a JCA.
The Commission selects devices for JCA following a recommendation from the HTA Coordination Group taking into account six criteria established under Article 7(4) of the HTA Regulation:
- unmet medical needs
- whether the technology is first in class
- potential impact on patients, public health or healthcare systems
- the incorporation of AI or machine-learning technologies
- its cross-border dimension
- its potential Union-wide added value.
To be selected, a MD or IVD needs to meet at least one of these six selection criteria. Selection takes place quarterly.
What is the process to select a device (MD or IVD) for JCA?
The selection process involves the following steps:
HTA Coordination Group recommendation
On a quarterly basis, the European Medicines Agency shares with the HTA Secretariat the list of expert panel opinions and views from the three previous months. The HTA Coordination Group may recommend MDs and IVDs from this list, taking into consideration the selection criteria set out in the HTA Regulation.
Information to the health technology developers and notified bodies
Once the HTA Coordination Group recommendation is adopted, the HTA Secretariat informs all concerned health technology developers and notified bodies whether their device has been recommended or not for a JCA.
In addition, for recommended devices, notified bodies are asked to inform the HTA Secretariat about the device’s certification and developers are invited to share the documentation relevant to the development of the JCA assessment scope, namely the certificate(s) of conformity and instructions for use.
In the letter informing the health technology developer that its device has been recommended for selection, the HTA Secretariat also provides the developer with technical instructions on how to access the secure HTA IT platform where all subsequent exchanges of information will take place.
To be noted that at this stage, information about which device(s) has been recommended for a JCA is not made public.
Selection by the Commission of the device for joint clinical assessment
Only once the recommended device has been certified can the Commission adopt an implementing decision selecting the device for joint clinical assessment. Once adopted, the decision is made public in the JCA tracker for medical devices.
How is a JCA on a device (MD or IVD) carried out?
Appointment of an assessor and co-assessor and selection of individual experts
Once a recommended device is certified, the JCA Subgroup appoints an assessor and co-assessor from two different Member States. The health technology developer is then informed of the names of the member organisations providing the assessor and co-assessor and given indicative dates for the next steps of the process.
The JCA Subgroup also selects individual experts (patients, clinicians and other relevant experts) to be consulted in this JCA.
Development of the assessment scope
Once the assessor and co-assessor are appointed, the JCA Subgroup starts developing the assessment scope of the joint clinical assessment. The scope defines the parameters to be assessed, including:
- the patient population
- the intervention
- the comparator(s)
the health outcomes.
The assessment scope reflects Member States’ needs in terms of these parameters. Selected patients, clinical experts and other relevant experts are consulted during the preparation of the assessment scope.
Once the Commission adopts the implementing decision selecting the device for JCA, the JCA Subgroup finalises the assessment scope for that device’s JCA.
Submission of the dossier
The Commission sends the finalised assessment scope to the health technology developer and requests submission of the dossier for the JCA (Commission’s first request).
The HTA Secretariat can organise an assessment scope explanation meeting with the JCA Subgroup at the request of the health technology developer.
The developer has 100 days to deliver the dossier. The dossier must be submitted digitally via the HTA IT Platform using the relevant template established in Annex I or II of Commission Implementing Regulation (EU) 2025/2086. The health technology developer shall consult, if available, the methodological guidance adopted by the HTA Coordination Group and other relevant documentation.
Commission’s confirmation check of the dossier
Once the dossier is submitted, the Commission, as appropriate in consultation with the assessor and co-assessor, checks whether it is submitted in a timely manner and meets the requirements of Article 9(2), (3) and (4) of the HTA Regulation.
Within 15 working days, the Commission either confirms the dossier’s compliance or sends the developer a second request with a list of missing items. In this case, the developer has 15 days to address that request and submit an updated dossier.
Preparation of the JCA report and summary report
Once the dossier is confirmed, the assessor and co-assessor begin preparing the JCA report. If needed, they may request additional specifications, clarifications, or other information from the developer.
Selected patients, clinical experts, and other relevant experts are then given the opportunity to provide feedback on the draft JCA report.
The JCA Subgroup is consulted on the draft report and any scientific issues are discussed within the Subgroup. Before the report is finalised by the JCA Subgroup, it is shared with the health technology developer for a factual accuracy check.
At the same time, the health technology developer may indicate any information it considers confidential, if applicable.
Endorsement and publication of the JCA report
The JCA results in:
- a Joint Clinical Assessment Report
a summary report.
The reports describe the scientific analysis of the available clinical evidence. They do not contain a value judgement or conclusion on the overall clinical added value of the MD or IVD.
The HTA Coordination Group endorses the final reports and the Commission publishes them on the HTA websites together with the JCA dossier, subject to applicable rules on confidentiality and the protection of commercially sensitive information.
What happens after the JCA?
The JCA report supports national HTA processes. Member States remain responsible for drawing conclusions about health technologies in their national HTA and for decisions concerning the use, pricing and reimbursement of health technologies.
Related information
- Implementing Regulation (EU) 2025/2086 on joint clinical assessment of medical devices and in vitro diagnostic medical devices
- Guidance on the HTACG recommendation for the selection of medical devices and in vitro diagnostic medical devices for JCA
- Procedural Guidance for JCA medical devices and in vitro diagnostic medical devices
- Guidance on filling in the joint clinical assessment (JCA) dossier template – Medical devices and in vitro diagnostic medical devices
- JCA tracker for medical devices and in vitro diagnostic medical devices